Choosing the right TFF filter for sample type and volume
"How much can it process?" is the first question anyone asks about a tangential flow filtration (TFF) system. The number on the datasheet is a ceiling, not a specification. A litre of clarified conditioned medium and a litre of raw urine are not the same litre. One runs to completion in ninety minutes on a 94 cm² cartridge; the other can foul the same cartridge before the volume is halved. Capacity in TFF is set by how much membrane area the feed demands.
This article works through what actually governs throughput, how each class of sample behaves, from mammalian conditioned medium to gold colloids, and how membrane material and format shift the answer. Where it helps, the numbers are given for the EVlution™ system, which runs polyethersulfone hollow fibre cartridges in two areas and ten pore sizes.
What Cartridge Capacity Actually Means
Four quantities decide whether a run finishes in an hour or stalls before completion.
Membrane area and volumetric loading. The useful unit is litres of feed per square metre of membrane, L/m². Bioprocess practice puts ultrafiltration loadings somewhere around 50 to 200 L/m² depending on how "dirty" the feed is. A 94 cm² cartridge is 0.0094 m². Running 1 L across it is a loading of about 106 L/m². A 260 cm² cartridge taking 3 L sits at about 115 L/m². Those are the same number, which is why the two EVlution cartridge sizes have the volume limits they do: not an arbitrary cut-off, but the same loading applied to different areas.
Permeate flux. Flux, in litres per square metre per hour (LMH), sets the clock. Ultrafiltration of a clean protein solution might run at 40 to 100 LMH; a fouling feed on a microfiltration membrane often stabilises between 15 and 100 LMH. For a 1 L sample concentrated to 20 mL on 94 cm², the permeate volume is roughly 0.98 L, and finishing in 90 minutes implies an average flux near 70 LMH. That is consistent with our in-house testing, where 1 L of cancer cell conditioned medium was concentrated to 20 mL in under 90 minutes at 1 bar.
Hold-up volume. Every system has liquid that can't elute without a wash. On EVlution fitted with 3 mm diameter tubing, the residual in the fluid path is around 10-20 mL, depending on configuration, and the void volume in hands-on Switch Flow™ mode is 5 mL. This is why sample volumes below 50 mL are not recommended: at 30 mL of feed, a third of the sample sits in the tubing before filtration starts.
Concentration factor and final volume. Minimising the final sample volume can be important in downstream processing and applications. For example, if EVs need further SEC clean-up, volume adds additional column cost. Switch Flow™ addresses this by driving the retentate back and forth through the cartridge with syringes once the loop volume reaches about 20 mL, bringing the recoverable sample to roughly 5 to 10 mL. A 3 L starting volume taken to 5 mL is a 600-fold concentration in a single unit operation.
| EVlution™ capacity by cartridge area | |||
| Parameter | 94 cm² | 260 cm² | Note |
| Maximum feed volume | 1 L | 3 L | Clean feeds only; fouling feeds need more area or less volume |
| Recommended working range | 100 mL to 1 L | 250 mL to 3 L | Below 50 mL is not recommended on either |
| Loading at maximum volume | ~106 L/m² | ~115 L/m² | The ceiling is the same on both; area is what changes |
| Typical final volume | <20 mL, or ~5 to 10 mL with Switch Flow™ | <20 mL, or ~5 to 10 mL with Switch Flow™ | Set by hold-up volume and tubing bore, not by cartridge area |
| Pore sizes available | 5 to 1,000 kDa (ten ratings) | 5 to 1,000 kDa (ten ratings) | Polyethersulfone hollow fibre throughout |
The Feed Decides The Capacity
Two feeds of identical volume can differ tenfold in the area they need. What matters is the load of material that will not pass the membrane: cells and debris, protein, lipid, polymer, salt at concentrations that affect viscosity. The table below sets out the main sample classes seen at laboratory and pilot scale, with the practical constraint for each.
| Sample types, typical volumes and what limits throughput | |||
| Sample | Typical volume | What limits capacity | Preparation that improves performance |
| Serum-free conditioned medium | 250 mL to 5 L | Cell debris and secreted protein | 300×g 5 min, 2,000×g 10 min, then 0.22 or 0.45 µm |
| Serum-containing medium, FBS-EV depletion | 100 mL to 1 L | Very high protein load; rapid polarisation | Dilute the serum before the run; expect lower flux throughout |
| Bioreactor or suspension harvest | 500 mL to 10 L | Solids load, cell lysis during handling | Clarify promptly; split across sequential cartridge loads if needed |
| Urine | 50 mL to 1 L | Uromodulin polymerises as the retentate concentrates, particularly at low pH | Process fresh, hold pH neutral, avoid taking the retentate to the lowest volume |
| Plasma or serum | 100 mL to 1 L after dilution | Viscosity and total protein; lipoproteins overlap the target size range | Dilute before loading; the dilution is what makes the volume figure look large |
| Saliva, CSF, lavage, ascites | 10 mL to 250 mL | Hold-up volume relative to sample; mucins in saliva | Shorten tubing; pool samples rather than running each one separately |
| Milk and cheesemaking whey | 100 mL to 2 L | Casein micelles and fat globules foul aggressively | Defat, acidify or treat with EDTA; whey loads far better than whole milk |
| Bacterial culture supernatant, OMVs | 250 mL to 5 L | Residual cells, exopolysaccharide | Full clarification; exopolysaccharide is the usual cause of unrecoverable fouling |
| Plant and microalgal material | 100 mL to 5 L | Cell wall fragments, pigments, phenolics | Coarse prefiltration; microalgal biomass can be returned to culture |
| Viral vectors and phage lysate | 250 mL to 5 L | Adsorptive loss and shear-sensitive capsids, not flux | Minimise passes and hold time; keep transmembrane pressure low |
| In vitro transcription and plasmid preparations | 50 mL to 2 L | Mass loading per area, and viscosity at high nucleic acid concentration | Work to a mass target per m², not a volume target |
| Recombinant protein and antibody solutions | 250 mL to 5 L | Final viscosity, which rises steeply above roughly 100 g/L | 0.22 µm filter; set a realistic end concentration before starting |
| Synthetic nanoparticles and colloids | 50 mL to 2 L | Surfactant and free polymer, aggregation at high concentration factor | Diafiltration volumes rather than concentration; keep the particles charged |
| Liposomes and lipid nanoparticles | 50 mL to 1 L | Residual ethanol reduces membrane permeability | Dilute the solvent before diafiltration rather than washing it out at full strength |
Biological Samples In Detail
Conditioned media. This is the best-behaved feed in the list and the one against which most capacity figures are quoted by us and other manufacturers. Comparative work on cancer cell conditioned medium found TFF gave higher yield than ultracentrifugation, fewer sub-15 nm aggregates, better batch-to-batch consistency and half the processing time.1 Serum-containing medium is a different proposition: the protein load is an order of magnitude higher and the flux reflects it. The same applies when the goal is to deplete FBS of its vesicles rather than to recover a product, where the material being rejected is the problem rather than the prize.
Biofluids. Urine is dilute, which makes it attractive for large starting volumes, but uromodulin is a filament-forming protein that co-concentrates with vesicles and can gel the retentate in the last few millilitres. Plasma and serum are the opposite: small volumes of a crowded, viscous matrix, almost always diluted before loading, which means a stated capacity of 1 L may represent 200 mL of plasma. Milk is the most demanding of the common biofluids. A recent head-to-head comparison of 300 kDa and 750 kDa membranes for bovine milk vesicles found both recovered particles well, but only the 750 kDa cartridge cleared the abundant milk proteins, β-casein and α-lactalbumin, effectively.2 Work on dairy whey reached the same conclusion from the other direction: a 750 kDa ultrafiltration membrane kept vesicle losses near zero but removed impurities less effectively than a 0.15 µm microfiltration membrane.3 The pore size chosen for a protein-rich feed is a purity decision as much as a capacity one.
Viruses and vectors. Here capacity is rarely about flux. Tandem TFF of lentiviral supernatant achieved more than 2,000-fold concentration in under three hours with recovery above 97%, using a 615 cm² 500 kDa module followed by a 40 cm² module of the same rating.4 The second cartridge exists purely to beat the hold-up volume of the first, which is the clearest illustration available of why final volume and feed volume are separate design problems. Earlier work on upscaled lentiviral production reported 66-fold concentration with complete recovery from 2 L of culture medium.5 The limiting factor in vector work is usually adsorption to the membrane and cumulative shear from repeated pump passes, not throughput.
Nucleic acids. Messenger RNA from in vitro transcription is the best-characterised recent case. Separation of mRNA from unreacted nucleoside triphosphates worked at a capacity load of about 19 g/m² with transmembrane pressure below 2.5 psi and a feed flux of 300 LMH, giving recovery above 70% with no detectable degradation; two sequential membrane washes cut losses from roughly 30% to about 3%.6 Note the unit: grams per square metre, not litres. For a dilute, high-value, strongly adsorbing solute, mass loading is the real constraint and volume is incidental.
Proteins, antibodies and peptides. The guidance for pore size is a cut-off three to six times below the molecular weight of the target, so a 30 kDa cartridge for a 150 kDa antibody. Capacity at the start of a protein run is set by area; capacity at the end is set by viscosity. Concentrating an IgG1 towards 200 mg/mL takes the solution past 20 cP, and diafiltration into a proline-containing buffer has been used to bring viscosity down from about 21.9 cP to about 11 cP. In-house protein testing on EVlution is more modest and more typical of laboratory work: 500 mL of 1 mg/mL BSA concentrated tenfold to 50 mL, repeated three times under the same conditions. Peptides sit at the other end, where a 5 or 8 kDa cartridge is doing desalting rather than separation, and the question becomes how many diafiltration volumes are needed rather than how much feed the cartridge will take.
Chemical And Non-Biological Samples
Diafiltration of water-soluble gold nanoparticles was shown twenty years ago to give rapid purification and size separation, removing small-molecule impurities and resolving particles of different core diameter.7 Polymeric nanoparticle work makes the capacity argument: poly(lactide-co-glycolide) particles purified of surplus polyvinyl alcohol by TFF reached 90.7% removal, and the time to remove 90% of the surfactant from a 200 mL sample was 2.8 hours by TFF against 9.6 hours by a centrifugal diafiltration device.8
Three things change when the sample is chemical rather than biological.
- Solvent compatibility becomes the first question. Polyethersulfone tolerates aqueous buffers, dilute acid and alkali, and ethanol, which is why EVlution cleaning uses 0.1 to 0.5 M NaOH and sterilisation uses 70% ethanol. Strong organic solvents are a different matter, and the instrument should not be used with corrosive or flammable liquids outside that cleaning step. Where a process genuinely needs dimethyl sulfoxide or similar, regenerated cellulose membranes are the usual choice.
- Solvent changes the membrane as well as the sample. Ethanol in liposomal feeds reduces hollow fibre permeability in inverse proportion to its concentration, with little consequence for a batch process but considerable consequence for continuous diafiltration; diluting the ethanol before diafiltration cuts the number of stages required.9 A flux figure measured in buffer is not transferable to a feed containing 20% ethanol.
- Stability, not retention, sets the end point. Colloids held at high concentration against a membrane aggregate. Diafiltration at constant volume is usually safer than concentration followed by washing, and the retentate should not be driven below the concentration at which the suspension is stable.
Polysaccharides and oligomers are an underused application. Fractionation of a macroalgal hydrolysate found 100 kDa sufficient for clarification with total retention of colloidal material, with oligosaccharides retained in the order glucans, then galactans, then arabinans, while peptides passed even a 5 kDa membrane; raising temperature from 20 to 50 °C and feed flow from 6.6 to 11.2 L/h improved flux without changing retention.10 Temperature is a throughput lever that biological samples rarely allow.
Choosing A Pore Size
| EVlution™ cartridge ratings and their applications | |||
| Rating | Approx. pore diameter | Retained | Typical use |
| 5 and 8 kDa | ~3 to 4 nm | Small proteins, peptides, nucleic acids | Desalting, removal of small molecules and residual solvent |
| 10 and 30 kDa | ~5 to 8 nm | Cytokines, growth factors, antibody fragments | Protein concentration and buffer exchange; the most common non-EV use |
| 50 and 100 kDa | ~12 to 18 nm | Albumin, large complexes, mRNA | Albumin removal, nucleic acid concentration, pre-filtration |
| 300 and 500 kDa | ~25 to 45 nm | Small EVs, viral vectors, most nanoparticles | The default for vesicle and vector work; passes free protein |
| 750 and 1,000 kDa | ~65 to 100 nm | Larger vesicles, microvesicles, apoptotic bodies | Protein-rich feeds such as milk; upper gate in a size-fractionation scheme |
Rejection rises with particle size as a probability rather than switching at a threshold, which the gold nanoparticle data in a systematic membrane study illustrate neatly: across a range of ultrafiltration membranes, 100 nm particles were held back almost completely, at 95 to 97%, while 50 nm particles of the same chemistry were rejected only around 43 to 63%.11 We have looked at the consequences of that in more detail in an article on gating EVs by size with sequential TFF.
Membrane Material
Material is the variable most often ignored and the one that determines how much of the sample arrives at the other end. The same study compared polyethersulfone, regenerated cellulose and highly cross-linked regenerated cellulose for adenovirus concentration.11 Polyethersulfone cassettes gave much the highest clean water permeability, around 880 to 925 LMH/bar, against roughly 560 for regenerated cellulose and 400 for the cross-linked variant. They also removed more impurity, clearing 86 to 91% of host cell protein and around 85% of DNA, partly by adsorption rather than sieving. The price was paid in fouling: water flux recovery after use was nearly three times better for the cellulose membranes, and cleaning with 1 M sodium hydroxide restored flux on most devices but not all. That is the trade in one paragraph. Hydrophobic membranes move more liquid and bind more material.
In practice the distinction is less stark than the polymer names suggest, because almost all current polyethersulfone ultrafiltration membranes are surface-modified to reduce protein binding, and alkaline cleaning recovers most of what fouling costs. Across the same study, a commercial 750 kDa polyethersulfone hollow fibre recovered 100% of infectious particles, while a cross-linked cellulose cassette rated near 300 kDa recovered 72%. Rating and format mattered more than polymer.
| Membrane materials used in laboratory and pilot TFF | |||
| Material | Character | Strengths | Watch for |
| Polyethersulfone, usually hydrophilised | Hydrophobic backbone, surface modified | High permeability, wide pore range, tolerates alkali and ethanol, reusable | Adsorption of hydrophobic solutes; needs disciplined cleaning |
| Polysulfone | Similar chemistry, older format | Robust, widely available | Generally lower flux than modified polyethersulfone |
| Regenerated cellulose and cross-linked forms | Hydrophilic | Low non-specific binding, narrow pore distribution, good solvent range including DMSO | Lower permeability; fewer large cut-offs available |
| Polyvinylidene fluoride | Hydrophobic, chemically resistant | Strong solvent and oxidant tolerance; favourable in some vector work | Mostly microfiltration ratings; limited ultrafiltration options |
| Ceramic | Inorganic | Extremes of temperature, pH and solvent; long service life | Cost and hold-up volume rule it out at laboratory scale |
Format matters alongside material. Hollow fibres present an open cylindrical lumen with laminar flow and no turbulence-promoting screen; flat sheet cassettes generate turbulence that raises flux but also raises local shear. The received wisdom that hollow fibres are gentler is now contested, since shear rate depends on flow and channel dimensions rather than on format alone. What is not contested is that an open lumen handles solids better, which is why hollow fibres dominate work on lysates, harvests and anything that has not been through a fine filter. We compare the formats in more depth in our introduction to TFF for EV research.
Pump Capacity And Tubing Diameter
Membrane area decides how much feed a cartridge can take. The pump decides how fast that feed crosses the membrane, and crossflow is what keeps the membrane surface swept. A cartridge sized correctly but fed too slowly fouls anyway.
EVlution uses a stepper-motor peristaltic pump running from a 24 V supply at under 48 W, with dosing accuracy better than ±2% and a quoted motor life of 6,000 hours. The head is fitted with 3.0 mm bore tubing (5.0 mm outside diameter) delivering at least 86 mL/min, and that is the configuration every protocol in the user guide assumes. The same pump serves the manual kits and EVlution Pro alike, so a method developed on one transfers to the other without re-establishing the crossflow.
Peristaltic flow is set by the bore of the tube in the head as much as by motor speed, since each revolution displaces the volume of tube swept by the rollers. What that flow then does at the membrane depends on the cartridge, and here the useful figure is not flow per unit area but wall shear rate in the fibre lumen:
γ = 4Qfibre / πR³
where Qfibre is the total flow divided by the number of fibres and R is the fibre inner radius. Membrane area does not appear. This is the point at which hollow fibre intuition parts company with flat sheet intuition: a screened cassette channel is short and thin and needs a high feed flow per unit area to generate turbulence, whereas a hollow fibre lumen is long, open and laminar, and is swept perfectly well at flows that would look inadequate expressed per square metre.
Both EVlution cartridges use 1 mm bore fibres in a 13 mm shell. The 94 cm² cartridge has an effective length of 15 cm and the 260 cm² cartridge 30 cm, which means most of the extra area in the larger cartridge comes from longer fibres rather than from more of them. That distinction matters, because length does not divide the flow and fibre count does.
| Crossflow conditions in the two EVlution™ cartridges at maximum pump output | |||
| Parameter | 94 cm² | 260 cm² | Note |
| Fibre bore and effective length | 1 mm, 15 cm | 1 mm, 30 cm | Same 13 mm shell in both |
| Approximate fibre count | ~20 | ~28 | 2.8× the area, but only 1.4× the fibres |
| Flow per fibre | ~4.3 mL/min | ~3.1 mL/min | Total flow divided by fibre count |
| Wall shear rate | ~730 s−¹ | ~520 s−¹ | Gentle end of the range used in EV and vector work |
| Axial pressure drop along the lumen | ~0.004 bar | ~0.006 bar | Negligible against a 1 bar TMP, so pressure is near uniform |
| Lumen hold-up volume | ~2.4 mL | ~6.6 mL | Part of the 10 to 20 mL system residual; the rest is tubing |
The larger cartridge therefore runs at roughly 70% of the shear of the smaller one at the same pump setting, not the 36% that a flow-per-area calculation would suggest. The ratio is simply the inverse of the fibre count. Anyone reaching for feed flux per square metre as a sizing criterion on a hollow fibre system will conclude that more flow is required than actually is.
Those figures are an upper bound rather than the working condition. 86 mL/min is the head's maximum, and the protocols run at 250 RPM, below it, so routine shear is lower again and there is flow in reserve for a feed that turns out to need more sweeping than expected.
Both figures sit at the gentle end of what the field uses. Reported hollow fibre work spans roughly 500 to 6,000 s−¹: adeno-associated virus has been harvested at 2,000 s−¹ and lentiviral vector concentrated at 5,000 to 6,000 s−¹, both with product intact, while a good deal of published TFF is run below 2,000 s−¹ as a matter of course. Operating at a few hundred per second is a deliberate position rather than a shortfall, and for lipid bilayer vesicles it is the right end of the range. The trade is that a heavily fouling feed receives correspondingly less sweeping, which is why clarification before the run matters as much on this system as the earlier sections suggest.
In practice the pump is rarely at maximum. The protocols specify 250 RPM for circulation, flushing and cleaning, dropping to 50 RPM for the transition into Switch Flow™ and to 25 RPM when refilling the fibre lumens with storage solution. Those reductions are there to keep air out of the fibres rather than to control flux. Pump speed sets crossflow; the flow control valve sets transmembrane pressure. Establish the first before touching the second.
Above 3 L. Past the benchtop range the module itself changes character. Larger cartridges gain area mostly by adding fibres rather than by lengthening them, so the feed flow that keeps lumen conditions comparable rises roughly in step with fibre count, and the plumbing grows with it. These are different classes of instrument built for different jobs rather than rungs of one ladder, but the progression is worth seeing in one place when a process is likely to be transferred later.
| Crossflow and plumbing from bench to production scale, at 1 mm fibre bore | |||
| Membrane area | Approximate fibre count | Feed flow at this scale | Typical tubing bore |
| 94 cm² | ~20 | Up to 86 mL/min | 3.0 mm |
| 260 cm² | ~28 | Up to 86 mL/min | 3.0 mm |
| 0.1 m² | ~105 | ~300 to 400 mL/min | 5 to 8 mm |
| 1 m² | ~530 | ~1.5 to 2 L/min | 10 mm or larger, often with manifolded modules |
Fibre bore is the other lever, and a stronger one than it looks. Shear goes as the inverse cube of the radius, so halving the bore at fixed area and fixed total flow raises wall shear roughly fourfold once the extra fibres needed to keep the area are accounted for. A system built on narrow fibres is inherently higher shear at the same pump setting than one built on wide fibres, which is worth knowing when comparing published shear figures between instruments. It also means fibre bore, not pump size, is usually the first thing to look at if an application needs more sweeping than it is getting.
Wider tubing is not free either. Line volume scales with the square of the bore, so the metre of fluid path that holds about 7 mL at 3 mm holds nearer 30 mL at 6.4 mm. Minimum recoverable volume therefore rises with scale at the same time as feed volume does, which is why pilot systems concentrate to hundreds of millilitres rather than to the 5 to 10 mL a benchtop system reaches with Switch Flow™. Getting a small final volume from a large starting volume ends up being a two-stage job, as the tandem lentiviral work discussed earlier did it.4
Pressure sets the ceiling on EVlution rather than flow. The tube connections can fail above roughly 1.5 bar, so 1 bar is the working pressure and 0.3 bar is specified for cleaning and rinsing. Larger systems use rigid sanitary fittings precisely because a push-fit connection will not hold at litres per minute. Scaling a TFF process past a few litres is a pump and plumbing problem more than a membrane one: the chemistry and pore rating carry across unchanged, and the hardware around them does not.
Two housekeeping points follow from the pump being peristaltic. The head tubing is a consumable: it fatigues where the rollers compress it, delivered flow drifts downwards as that happens, and a tube that splits under pressure sprays sample. Replace it at the interval the manufacturer specifies rather than when it fails. Calibration is by timed delivery into a measuring cylinder, and is worth repeating whenever a new cartridge is fitted. Note too that the volume on the pump display counts liquid that has passed through the head, not sample processed. Because the loop recirculates, the same fluid crosses the head many times and the displayed figure runs well ahead of the feed volume.
Lab Scale, Pilot Scale And What Changes Between Them
Scaling TFF is done by holding loading and flux constant and varying area, which is why the figures given earlier matter more than the volume limits. A process characterised at 100 L/m² and 40 LMH on 94 cm² should behave the same on 260 cm², and the same again on a square metre of the same membrane chemistry, provided the fibre dimensions and therefore the shear rate are comparable. Hollow fibre geometry makes this relatively honest, since the lumen diameter and length can be matched across module sizes. The part that does not scale by itself is the pump: holding flux constant while area grows means crossflow per unit area falls unless pump capacity grows with it, which is the first thing to check when moving a method past 3 L.
| Working bands from bench to pilot | |||
| Band | Volume | Indicative area | Dominant constraint |
| Micro | Under 50 mL | Tens of cm² | Hold-up volume dominates; TFF is usually the wrong tool |
| Bench | 50 mL to 1 L | ~94 cm² | Time per run; recovery at low final volume |
| Extended bench | 1 to 3 L | ~260 cm² | Fouling over a longer run; cartridge life |
| Pilot | 3 to 50 L | 0.1 to 1 m² | Buffer consumption, hold time, product stability during the run |
Two things that matter little at bench scale become visible at pilot scale. The first is buffer. Discontinuous diafiltration at a fivefold dilution per cycle takes a 20 mL retentate to better than 99% exchange in three cycles using 240 mL of buffer; the same exchange in single-volume steps takes seven cycles and 140 mL. Which is cheaper depends on scale and on how well the product tolerates dilution. The second is time. A run that takes 90 minutes at 1 L takes most of a day at 30 L unless area scales with it, and hold time at room temperature is itself a stability variable for vesicles and vectors.
Where Capacity Is Lost
Most disappointing runs fail for one of four reasons, none of which is the headline volume.
- Inadequate clarification. Debris and aggregates that would have taken thirty minutes to remove by centrifugation and dead-end filtration instead occupy the cartridge for its remaining life. This is the single biggest determinant of how many of the twenty permitted reuses a cartridge actually delivers.
- Pressure instead of crossflow. Crossflow is established first at 250 RPM, and only then is the valve closed to generate transmembrane pressure. Reaching for pressure to fix a slow permeate rate compacts the fouling layer and makes it worse, which is why cleaning is specified at 0.3 bar rather than the 1 bar used for processing. The mechanism is covered in our article on flux and pressure during TFF.
- Adsorption. Membrane binding is invisible in a flux trace and shows up only in a mass balance. Measuring the target in the feed, the retentate and the permeate is the only way to tell a sieving loss from a binding loss, and the remedies differ: a different pore size for the first, a wash step or a passivated membrane for the second.
- Air. Drawing air into the fibres generates surface tension forces that will denature protein and rupture vesicles, and can airlock sections of membrane permanently. The pump stops when the retentate falls below 10 mL, not after.
Standardization
There is no agreed way to specify TFF capacity for a given feed. Manufacturers quote volumes, the literature reports loadings in L/m² or g/m², and the two are not interconvertible without knowing the solids content of the sample. A standardised fouling index for common biological feeds would let laboratories size a cartridge from a 10 mL screening run rather than by trial. Nothing of the kind exists.
Nominal ratings are also not comparable between manufacturers, so a 500 kDa cartridge from one supplier may sit between the 300 and 500 kDa cartridges of another. Until ratings are tied to a measured sieving curve rather than a dextran convention, the only reliable approach is to characterise the fractions produced by a given cartridge on a given feed, and to treat the number on the label as a starting point for a design rather than an answer.
References
- Busatto S, Vilanilam G, Ticer T, et al. Tangential flow filtration for highly efficient concentration of extracellular vesicles from large volumes of fluid. Cells. 2018;7(12):273.
- Zhang L, Xu R, Wang F, et al. Comparative profiling of tangential flow filtration and ultracentrifugation for the scalable production of bovine milk extracellular vesicles. Syst Microbiol Biomanuf. 2026. doi:10.1007/s43393-026-00499-9.
- Onesti R, Giancaterino S, Masciolini L, Piunti C, Cimetta E, Boi C. Continuous tangential flow filtration to isolate extracellular vesicles from dairy by-products. Ind Eng Chem Res. 2025.
- Cooper AR, Patel S, Senadheera S, Plath K, Kohn DB, Hollis RP. Highly efficient large-scale lentiviral vector concentration by tandem tangential flow filtration. J Virol Methods. 2011;177(1):1-9.
- Geraerts M, Michiels M, Baekelandt V, Debyser Z, Gijsbers R. Upscaling of lentiviral vector production by tangential flow filtration. J Gene Med. 2005;7(10):1299-1310.
- Nourafkan E, Kenyon C, Nair A, et al. An experimental and modeling approach to study tangential flow filtration performance for mRNA drug substance purification. Biotechnol J. 2024;19(11).
- Sweeney SF, Woehrle GH, Hutchison JE. Rapid purification and size separation of gold nanoparticles via diafiltration. J Am Chem Soc. 2006;128(10):3190-3197.
- Dalwadi G, Benson HAE, Chen Y. Comparison of diafiltration and tangential flow filtration for purification of nanoparticle suspensions. Pharm Res. 2005;22(12):2152-2162.
- Worsham RD, Thomas V, Farid SS. Impact of ethanol on continuous inline diafiltration of liposomal drug products. Biotechnol J. 2023. doi:10.1002/biot.202300194.
- Trigueros E, Sanz MT, Beltrán S, Ruiz MO. Filtration of subcritical water hydrolysates from red macroalgae byproducts with ultraporous ceramic membranes for oligosaccharide and peptide fractionation. J Membr Sci. 2022;660:120822.
- Nestola P, Martins DL, Peixoto C, et al. Evaluation of novel large cut-off ultrafiltration membranes for adenovirus serotype 5 (Ad5) concentration. PLoS One. 2014;9(12):e115802.
The EVlution™ TFF system runs polyethersulfone hollow fibre cartridges in two membrane areas, 94 cm² for samples up to 1 L and 260 cm² for samples up to 3 L, across ten pore sizes from 5 kDa to 1,000 kDa. The same stand, pump and fluid path serve the whole range, so a method developed on a protein at 10 kDa and a method developed on vesicles at 500 kDa differ only in the cartridge fitted. Full operating parameters, cleaning and storage procedures are in the EVlution User Guide.