Exo-spin maxi SEC compliments EV TFF

Exo-spin maxi SEC compliments EV TFF

Exo-spin™ columns have been used for over 10 years to purify exosomes and other extracellular vesicles (EVs) and reported in more than 1000 peer-reviewed papers, mostly at small scale. Until now the largest column in the range was the midi, which has a 10 ml bed and takes 1 ml of plasma or serum directly. For biomarker work that is usually plenty.

When the experiment downstream needs a lot of material a bigger column is needed. Applications requireing large amounts of EVs include dosing in animal models, functional assays run in replicate, or a reference batch of EVs that has to last a whole project. The new Exo-spin™ maxi (EX08-1) has a 70 ml bed and takes up to 5 ml of plasma or serum or a TFF-concentrated sample in a single load.

How the maxi works

The principle is the same as the rest of the Exo-spin™ range. The column is packed with a porous size exclusion resin with pores of about 30 nm. Proteins and other small molecules enter the pores and take a longer path through the bed, while EVs are too large to get in and travel around the beads, so they come off the column first. Exosomes in the 30 to 200 nm range elute in the early fractions, and around 99.5% of the other proteins and lipids are held back and elute later.

Maxi columns come packed, equilibrated and ready to use. The run is by gravity flow with fractions collected as the column drips, takes under two hours, and needs no ultracentrifuge. Each column can be reused up to five times.

Where the maxi sits in the Exo-spin™ range

 

Exo-spin™ column sizes compared
Column Bed volume Direct load (plasma or serum) Typical use
mini (EX01, EX02, EX03) 0.5 ml 100 µl Fast, low-cost isolation from small samples
mini-HD (EX05) 3.5 ml 150 µl Higher purity and finer fractions from small samples
midi (EX04) 10 ml 1 ml Larger samples, up to 500 ml media with precipitation
maxi (EX08) 70 ml 5 ml Preparative scale, litres of media after TFF

 

For plates of samples there is also the Exo-spin™ 96, which is the opposite end of the problem: many small samples in parallel rather than one big one.

Why the bed size matters

How well a size exclusion column separates EVs from protein depends mainly on two things; the length of the bed and how big the sample is compared with it. A 5 ml load on a 70 ml bed is about 7% of the bed volume. That is in the same region as the 1 ml on a 10 ml Sepharose CL-2B column used by Böing and colleagues in the paper that did much to popularise SEC for plasma EVs, where vesicles were well separated from both protein and HDL.

In our own test, 5 ml of 1% FBS was loaded directly onto a maxi column and 1 ml fractions were collected. Nanoparticle tracking analysis put the EVs in fraction 20, and the BCA assay showed the first protein coming through at fraction 25. Five fractions is a comfortable gap. It leaves room to pool the EV fractions without cutting into the protein front, which is harder to do on short columns where the two peaks sit closer together.

Pairing the maxi with tangential flow filtration

Size exclusion has two built-in limits. It dilutes the sample, and it has a fixed maximum load. Plasma is rich in EVs, so 5 ml loaded straight on is a good amount of material. Culture media, urine and saliva are not, and 5 ml of conditioned media on its own gives very little back.

The answer is to concentrate first. Benedikter and colleagues showed that ultrafiltration followed by SEC recovers EVs from culture media at least as well as ultracentrifugation, with better purity, and the approach has become a standard route. The maxi was sized with this in mind: 5 ml is a volume that a TFF system reaches easily from a large starting volume. In one of our runs, 500 ml of LNCaP conditioned media was concentrated to 5 ml on an EVlution™ TFF system and loaded onto a maxi column, and western blotting found the EV marker ALIX in fractions 17 to 20. Media volumes up to about 3 litres can be handled the same way.

The two steps divide the work well. TFF takes the volume down and, depending on the cartridge pore size, washes out a good part of the free protein on the way. The column then does the fine separation. For smaller volumes of dilute sample, precipitation with Exo-spin™ buffer (EX06) remains an option, but skipping the concentration step altogether will lower recovery.

Some practical points

Run a profile the first time you use the column with a new sample type. Collect 1 ml fractions across the whole run and check particles (NTA or similar) and protein in each, then decide which fractions to pool. The position of the EV peak can move a little with sample viscosity and composition, and plasma behaves differently from concentrated media.

The pooled EV fractions will add up to a few millilitres. If the downstream assay needs a small volume, a second concentration step is needed, either on a small TFF cartridge or a centrifugal concentrator.

SEC separates by size, so anything of a similar size to EVs comes off with them. In plasma this includes the larger lipoproteins. MISEV2023 discusses this at some length, and for some applications, lipidomics for example, a second orthogonal step may be worth adding. For many others it will not matter. It depends on what you are measuring.

Plasma and serum should be cleared of cells and platelets by centrifugation before loading, as with any of the Exo-spin™ columns.

References

  1. Böing AN, van der Pol E, Grootemaat AE, Coumans FAW, Sturk A, Nieuwland R. Single-step isolation of extracellular vesicles by size-exclusion chromatography. J Extracell Vesicles. 2014;3:23430.
  2. Benedikter BJ, Bouwman FG, Vajen T, et al. Ultrafiltration combined with size exclusion chromatography efficiently isolates extracellular vesicles from cell culture media for compositional and functional studies. Sci Rep. 2017;7:15297.
  3. Welsh JA, Goberdhan DCI, O'Driscoll L, et al. Minimal information for studies of extracellular vesicles (MISEV2023): From basic to advanced approaches. J Extracell Vesicles. 2024;13(2):e12404.

The Exo-spin™ maxi is available now alongside the rest of the Exo-spin™ range. Cell Guidance Systems also offers EV and exosome services, including isolation and NTA size profiling, if you would rather we did the purification for you.